background image

Calibration Information

All gauges are factory calibrated on NIST traceable 
calibration equipment. No calibration is required before 
placing the gauge into service.
Gauges can be returned to Omega Engineering for factory 
certified recalibration. NIST traceability is available. 
Gauges can also be recalibrated by any metrology lab with 
pressure calibration equipment at least four times more 
accurate than the gauge.
Absolute reference models display atmospheric pressure if 
the gauge port is open to the ambient. It is normal for the 
reading to constantly change in response to atmospheric 
pressure changes. Vacuum generation and atmospheric 
pressure measurement equipment for accurate calibration 
and thus these are more difficult to calibrate in the field.

Calibration Preparation – All Models

All gauges are factory calibrated using NIST traceable 
calibration equipment. Calibration is not required before 
using the gauge. Calibration intervals depend on your 
quality standards, but annual re-calibration is customary.
Calibration equipment is not required to zero gauge 
reference ranges. Absolute reference ranges may be 
zeroed with application of full vacuum. 
Span calibration should only be performed using 
appropriate calibration procedures with calibration 
standards that are at least four times more accurate than 
the gauge being calibrated.
The calibration system must be able to generate and 
measure pressure/vacuum over the full range of the gauge. 
A vacuum pump able to produce a vacuum of 100 microns 
(0.1 torr or 100 millitorr) or lower is required for vacuum 
and absolute gauges.
It is good practice to install fresh batteries before 
calibrating battery-powered gauges.
Allow the gauge to equalize to normal room temperature 
for approximately 20 minutes before calibration.

Calibration of Ranges up to 1999

For DPG1000AD and DPG1100AD models remove the 
front covers to access the zero and span calibration 
potentiometers. 
For DPG1100AD models unscrew the nylon cover screws to 
access the zero and span calibration potentiometers.
Gauges may be re-zeroed without affecting the span 
calibration. For gauge reference models the gauge port 
must be open to the ambient. For absolute reference 
models full vacuum must be applied. 
Adjust the zero control until the gauge reads zero with the 
minus (–) sign occasionally flashing.
Zero calibration must be done before span calibration. 
Using the appropriate pressure standards, record readings 
at three to five points over the range of gauge and adjust 
span control to minimize error and meet specifications.

Calibration of Ranges 2000 and up including 
3000 psi, 5000 psi

Entering Calibration Mode

Remove the rear cover to gain access to the UP and DOWN 
buttons located near the lower right and left corners of the 
circuit board.

With the gauge off, press and hold the DOWN calibration 
button, and also press the front button.
The full-scale pressure range and display test is shown, 
and then CAL is displayed to indicate that the calibration 
mode is enabled.
Release all buttons. The gauge enters and remains in 
the calibration mode until restarted manually or power is 
removed. Features not related to calibration are disabled. 
If the power is removed during calibration, settings will not 
be saved.
The display will indicate the current pressure reading, 
updating approximately 3 times per second. 
Each press of the UP or DOWN button makes a small 
correction, which may not always be indicated on the 
display. Press and hold the button for one second or longer 
to make larger corrections. The gauge’s display is adjusted 
to match the calibrator’s reading.

Gauge Reference Ranges (3 Points)

With the gauge port open to atmosphere, the character 
display will alternate between ZERO and CAL. Press the UP 
and DOWN buttons to obtain a display indication of zero.
Apply full-scale pressure (or vacuum for vacuum gauges). 
The character display will alternate bSPAN and 
CAL. Press the UP and DOWN buttons to obtain a display 
indication equal to full-scale pressure.
Apply 50% of full-scale pressure. The character display will 
alternate bMID and CAL. Use the UP and DOWN 
buttons to obtain a display indication equal to 50% of full-
scale pressure.

Absolute Reference Ranges (3 Points)

Apply full vacuum to the gauge. The character display will 
alternate between ZERO and CAL. Press the UP and DOWN 
buttons to obtain a display indication of zero.
Apply full-scale pressure. The character display will 
alternate bSPAN and CAL. Press the UP and 
DOWN buttons to obtain a display indication equal to full-
scale pressure.
Apply 50% of full-scale pressure. The character display will 
alternate bMID and CAL. Press the UP and DOWN 
buttons to obtain a display indication equal to 50% of full-
scale pressure.

Bipolar (±) and Compound Ranges (4 or 5 Points)

With the gauge port open to atmosphere, the character 
display will alternate between ZERO and CAL. Press the UP 
and DOWN buttons to obtain a display indication of zero.
Apply full-scale positive pressure. The character display 
will alternate bSPAN and CAL. Press the UP and 
DOWN buttons to obtain a display indication equal to full-
scale pressure.
Apply 50% of full-scale positive pressure. The character 
display will alternate bMID and CAL. Press the UP 
and DOWN buttons to obtain a display indication equal to 
50% of full-scale pressure.
Apply full vacuum. The character display will alternate 
between –SPAN and CAL. Press the UP and DOWN buttons 
to obtain a display indication equal to the full vacuum 
reading.
Gauges using a ±15 psig sensor have a –MID calibration 
point. Apply 50% of the full-scale vacuum range (for 
example, –7.4 psi for a ±15 psi gauge). The character 
display will alternate between –MID and CAL. Press the UP 
and DOWN buttons to obtain a display indication equal to 
50% of full-scale vacuum.

Exit Calibration Mode and Verify Calibration

Exit the calibration mode and save the calibration data 
by pressing and holding the front button until the display 
indicates OFF.
Verify readings at 0%, 25%, 50%, 75%, and 100% of full 
scale.
Replace the rear cover and screws, taking care not to 
pinch the wires between the case and the rear cover.

DPG1000B, DPG1001B, DPG1000BBL, DPG1100B, DPG1100BBL

3

NOTE:

DPG1000, DPG1001

DPG1100

DOWN

UP

DPG1100B, DPG1100BBL 

Ranges above 2000

DPG1000B, DPG1001B, 

DPG1000BBL Ranges above 2000

Summary of Contents for DPG1000B

Page 1: ... 2000 DPG1100B DPG1100BBL Ranges abovve 2000 PSI Res inHg PSI Res mmH2O Res 3PSIG 01 30V15PSIG 1 2000MMH2OG 1 5PSIG 01 30V100PSIG 1 3500MMH2OG 1 15PSIA 01 30V200PSIG 1 cmH2O Res 15PSIVAC 01 inH2O Res 200CMH2OG 1 15PSIG 1 85INH2OG 1 350CMH2OG 1 15PSIG 01 140INH2OG 1 1000CMH2OA 1 30PSIA 1 400INH2OA 1 1000CMH2OVAC 1 30PSIG 1 400INH2OVAC 1 1000CMH2OG 1 60PSIG 1 400INH2OG 1 1000CMH2OG 1 100PSIA 1 400IN...

Page 2: ... The actual pressure is displayed Attempting to zero the gauge with pressure greater than approximately 3 of full scale applied will result in an error condition and the display will alternately indicate E r r 0 and the actual measured pressure The gauge must be powered down to reset the error condition Absolute reference gauges do not use the zero feature since they read atmospheric pressure unde...

Page 3: ...s in the calibration mode until restarted manually or power is removed Features not related to calibration are disabled If the power is removed during calibration settings will not be saved The display will indicate the current pressure reading updating approximately 3 times per second Each press of the UP or DOWN button makes a small correction which may not always be indicated on the display Pre...

Page 4: ...by OMEGA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to mishandling improper interfacing operation outside of design limits improper repair or unauthorized modification This WARRANTY is VOID if the unit shows evidence of having been tampered with or ...

Reviews: