background image

DPG1000B, DPG1001B, DPG1000BBL, DPG1100B, DPG1100BBL

2

Installation and Precautions

4

 Read these instructions before using the gauge. Contact 

the factory for assistance.

4

 These products do not contain user-serviceable parts. 

Contact us for repairs, service, or refurbishment.

4

 Gauges must be operated within specified ambient tem-

perature ranges.

4

 Outdoor or wash down applications require a NEMA 4X 

gauge or installation in a NEMA 4X housing.

4

 Use a pressure or vacuum range appropriate for the 

application.

4

 Use fittings appropriate for the pressure range of the 

gauge. 

4

 Due to the hardness of 316 stainless steel, it is recom-

mended that a thread sealant be used to ensure leak-
free operation.

4

 For contaminated media use an appropriate screen or 

filter to keep debris out of gauge port. 

4

 Remove system pressures before removing or installing 

gauge.

4

 Install or remove gauge using a wrench on the hex fitting 

only. Do not attempt to turn gauge by forcing the housing. 

4

 Good design practice dictates that positive displacement 

liquid pumps include protection devices to prevent sen-
sor damage from pressure spikes, acceleration head, 
and vacuum extremes.

Avoid permanent sensor damage! Do not apply vacuum 
to non-vacuum gauges or hydraulic vacuum to any 
gauges. 

Avoid permanent sensor damage! NEVER insert objects 
into gauge port or blow out with compressed air.

 Gauges are not for oxygen service. Accidental rupture of 

sensor diaphragm may cause silicone oil inside sensor to 
react with oxygen.

Absolute Reference Gauges

These models have engineering units shown on the rear 
label such as PSIA, INHGA, INH2OA, ZIN2A, BARA, MBARA, 
KPAA, TORR or TORRA, MMHGA, KGCMA, ATMA. The 
“A” indicates absolute reference. These ranges display 
atmospheric pressure when the gauge port is open to the 
atmosphere. They display zero at high vacuum. 
When the gauge port is open to the atmosphere, it is 
normal for the reading to constantly change in response 
to atmospheric pressure changes. Do not adjust the 
calibration controls!

Operation – Ranges up to 1999

Press the round button on the front of the gauge to activate 
the display. The gauge will stay on for a period of time 
determined by the auto-shutoff time. The gauge can be 
shut off at any time by pressing the button again. 
If the gauge was ordered without auto shutoff it will stay on 
until the button is pressed or until the batteries are depleted. 
Turn gauge off when not in use to conserve battery.
Display backlighting on BBL models is on whenever the 
gauge is on. Backlighting is not apparent under bright 
conditions.

Operation – Ranges above 2000 including 
3000 psi, 5000 psi

These ranges have a 4-digit display with a lower 
alphanumeric display. Press and hold the pushbutton for 
approximately 1 second. The full-scale range is indicated, 
display segments are tested, and the reading and units are 
displayed.

Power-Up With Zero – Ranges above 2000 
including 3000 psi, 5000 psi

1. Make absolutely certain no pressure is applied to the 

gauge. The gauge port should be exposed to normal 
atmospheric pressure. Note that the zeroing function 
may only be activated at power-up and the stored zero 
correction is erased when the gauge is shut off.

2. Press and hold the pushbutton.

3. The full-scale range is indicated and the display segments are 

tested. 

4. Continue to press the pushbutton until o o o o  is displayed 

and then release the button. This indicates that the gauge has 
been zeroed.

5. The actual pressure is displayed.

Attempting to zero the gauge with pressure greater than 
approximately 3% of full-scale applied will result in an error 
condition, and the display will alternately indicate 

E

 

r r

 

0

 

and the actual measured pressure. The gauge must be 
powered down to reset the error condition. 
Absolute reference gauges do not use the zero feature 
since they read atmospheric pressure under normal 
conditions.

Normal Operation – Ranges 2000 and up 
including 3000 psi, 5000 psi

Following the start-up initialization, the display indicates 
the pressure reading updated approximately 3 times per 
second and the units. The auto shutoff timer starts when 
the gauge is powered up or whenever the button is pushed, 
unless the gauge was ordered without an auto shutoff time 
(

-NS

 option). 

If excessive vacuum is applied to a pressure-only gauge, 
the display will indicate 

 

E

 

r r

  until the vacuum is 

released. Applying vacuum to a gauge designed for 
pressure may damage the pressure sensor. If excessive 
pressure is applied (112.5% over range), an out-of-
range indication of 

i

 – – – or 

i

 –.–.–.– will be displayed 

depending on model.

Display Backlighting (BBL models only)

Display backlighting can be turned on by momentarily 
pressing the button whenever the gauge is on. 
The backlighting will turn on for one minute and then 
automatically shut off. This also restarts the auto shutoff 
timer. Backlighting is not apparent under bright conditions.

Shut-Down

To shut off the gauge manually at any time, press and hold 
the pushbutton until the display indicates O F F (about 5 
seconds) and then release.
For gauges with auto shutoff, the display indicates O F 
F five seconds prior to auto shutoff. The pushbutton can 
be pressed to keep the gauge on. The auto shutoff and 
backlight (if equipped) timers are reset whenever the 
pushbutton is pressed and released.
If the gauge was ordered without auto shutoff, it will 
stay on until manually shut off or until the batteries are 
depleted. Turn gauge off when not in use to conserve 
battery life.

Battery Replacement

A low battery indication will be shown in the upper left-
hand corner of the display when the battery voltage falls 
sufficiently. The battery should be replaced soon after the 
indicator comes on or unreliable readings may result.
Remove the 6 Phillips head screws on the back of the unit. 
Remove the battery holder or cover depending on the 
model. 
Carefully remove batteries from the holders by lifting up the 
positive end of the battery (opposite the spring). Take care 
not to bend or distort the battery retention springs.
DO NOT discard the old battery into fire, any other sources 
of extreme heat, or in any other hazardous manner. Please 
consult local authorities if there is any question about 
proper disposal.
Always replace both batteries at the same time with high 
quality alkaline batteries. Observe the polarity of the 
batteries when replacing them. The negative (flat) end of 
each battery should be inserted first, and should face the 
spring in the battery holder.
Replace the back cover, including the rubber sealing 
gasket.

CAUTION:

Turn at 

hex 

fitting 

only

Turn at 

hex 

fitting 

only!

1

/

4

”NPT

3.38”

0.75”

PSIG

Span

Zero

2.88”

1.65”

1000B, 1001B, 1000BBL

OMEGA Engineering Inc./Made in USA

3

1

/

2

 digit or  

4

1

/

2

 digit display

1

/

4

”NPT

3.0”

3.5”

2.0”

PSIG

Span

Zero

OMEGA Engineering Inc./Made in USA

0.75”

1100B and 1100BBL (NEMA 4X)

3

1

/

2

 digit or  

4

1

/

2

 digit display

WARNING:

 This product can expose you to chemicals including 

nickel and chromium, which are known to the State of Califor-

nia to cause cancer or birth defects or other reproductive harm. 

For more information go to www.P65Warnings.ca.gov

Summary of Contents for DPG1000B

Page 1: ... 2000 DPG1100B DPG1100BBL Ranges abovve 2000 PSI Res inHg PSI Res mmH2O Res 3PSIG 01 30V15PSIG 1 2000MMH2OG 1 5PSIG 01 30V100PSIG 1 3500MMH2OG 1 15PSIA 01 30V200PSIG 1 cmH2O Res 15PSIVAC 01 inH2O Res 200CMH2OG 1 15PSIG 1 85INH2OG 1 350CMH2OG 1 15PSIG 01 140INH2OG 1 1000CMH2OA 1 30PSIA 1 400INH2OA 1 1000CMH2OVAC 1 30PSIG 1 400INH2OVAC 1 1000CMH2OG 1 60PSIG 1 400INH2OG 1 1000CMH2OG 1 100PSIA 1 400IN...

Page 2: ... The actual pressure is displayed Attempting to zero the gauge with pressure greater than approximately 3 of full scale applied will result in an error condition and the display will alternately indicate E r r 0 and the actual measured pressure The gauge must be powered down to reset the error condition Absolute reference gauges do not use the zero feature since they read atmospheric pressure unde...

Page 3: ...s in the calibration mode until restarted manually or power is removed Features not related to calibration are disabled If the power is removed during calibration settings will not be saved The display will indicate the current pressure reading updating approximately 3 times per second Each press of the UP or DOWN button makes a small correction which may not always be indicated on the display Pre...

Page 4: ...by OMEGA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to mishandling improper interfacing operation outside of design limits improper repair or unauthorized modification This WARRANTY is VOID if the unit shows evidence of having been tampered with or ...

Reviews: