Chapter 2
Maintaining a Split/Splitless Injector
Clean or Replace the Liner
34
Maintenance and Troubleshooting Manual
OPERATING SEQUENCE
Clean or Replace the Liner
Materials needed
•
Liner cap removal tool (P/N 205 070 10)
•
Glass liner (see Table 2-2, on page 29)
•
Graphite seal (P/N 290 334 06)
•
Ultrasonic bath
•
Methanol/acetone mixture (1:1)
•
Non metallic sharp tool
•
Septum (P/N 313 032 11)
•
Tweezers
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the injector must be cooled to room temperature.
PRECAUTIONS
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
for more information.
1. While in stand-by condition, press
OVEN
to
access the oven control table.
.
OVEN
Temp
50
50 <
Initial time
2
Ramp 1
10.0
2. Scroll to
Temp
and press
OFF
.