Chapter 15
Checking for Leaks
Maintaining a Photoionization Detector
Maintenance and Troubleshooting Manual
329
3. Press
LEFT CARRIER
or
RIGHT CARRIER
depending on which injector is operating.
XXX CARRIER (XXX)
Col. flow
2.5
<
Pressure
100
200
Flow mode
con pres
4. Scroll to
Flow mode
and set
con pres
.
5. Scroll to
Pressure
and set
200
.
6.
.
XXX DET (PID)
Lamp
Off <
Base temp
250
250
High current
N
Signal pA
(xx)
Mkup(N2)
30
Off
Sheath gas
40
200
Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the PID detector,
to access the PID control table.
7. Set
Lamp
to
Off
.
8. Set
Mkup(N2)
to
Off
.
9. Set
Sheath
gas to
200
.
10. Wait some minutes until the pressure in the
system is equilibrated.
11.
.
XXX DET (PID)
Lamp
Off <
Base temp
250
250
High current
N
Signal pA
(xx)
Mkup(N2)
30
Off
Sheath gas
40
Off
Set
Sheath gas
to
Off
.