Chapter 11
Maintaining a FID
Clean the Jet
196
Maintenance and Troubleshooting Manual
1.
.
RIGHT DETECTOR (FID)
Flame
Off <
Base temp
300
300
Signal pA
(5.4)
Ign. tresh
2.0
Flameout retry
Off
H2
35
Off
Air
350
Off
Mkup N2
30
30
Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the FID detector,
to access the FID control table. In the following
example, a FID installed on the right channel is
considered.
2. Scroll to
Flame
and press
OFF
.
3. Scroll subsequently to
Base temp,H2
and
Air
and press
OFF
.
4. Disconnect the signal and ignition polarization
cables from the detector.
5. Loosen the fixing screw on the front of the
detector cell and remove it.