Chapter 3
Maintaining a Large Volume Splitless Injector
Replace a Broken Liner
72
Maintenance and Troubleshooting Manual
21. Tighten the liner cap using the liner cap wrench
provided with the GC.
22. Place the septum holder with the septum onto
the inlet port, then screw and tighten the ring
nut up to the stop.
23. Tighten the injector cap to finger tight.
CAUTION
Do not overtighten the injector
cap. You could damage the
septum and affect performance.
24. Press
LEFT CARRIER
or
RIGHT CARRIER,
depending on which injector is operating, scroll
to
Pressure
and press
ON
.