Chapter 13
Replace the NPD Thermionic Source
Maintaining a Nitrogen Phosphorus Detector
Maintenance and Troubleshooting Manual
261
1. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the NPD detector,
to access the NPD control table. In the
following example, a NPD installed on the right
channel is considered.
.
RIGHT DETECTOR (NPD)
Source cur,A
Off <
Base temp
300
Off
Signal pA
(10.2)
Polarizer V
3.5
H2 delay time
Off
H2
2.3
Off
Air
70
Off
Mkup N2
15
Off
2. Scroll to
Base temp
and press
OFF
.
3. Scroll subsequently to
Source curr,H2
,
Air
and
Mkup
and press
OFF
.
4. Disconnect the signal and source cables from
the detector.
5. Loosen the fixing screw on the front of the
detector cell and remove it.
Содержание TRACE GC Ultra
Страница 4: ......
Страница 10: ...x Maintenance and Troubleshooting Manual Contents ...
Страница 20: ......
Страница 128: ......
Страница 178: ...Chapter 9 Maintaining a PTV Injector PTV Troubleshooting 178 Maintenance and Troubleshooting Manual ...
Страница 218: ......
Страница 268: ......
Страница 314: ......