Chapter 2
Maintaining a Split/Splitless Injector
Replace a Broken Liner
46
Maintenance and Troubleshooting Manual
23. Remove the septum from the septum holder
(use non-metallic tools).
24. Insert a new septum into the septum holder.
CAUTION
Use tweezers to avoid touching
the septum with your fingers.
25. Clean the septum support from possible
fragments left by the septum and reinsert it into
the injector
26. Place the septum holder on the top of the
septum support.
27. Tighten the injector cap to finger tight.
CAUTION
Do not overtighten the injector
cap. You could damage the
septum and affect performance.
28. Press
LEFT CARRIER
or
RIGHT CARRIER,
depending on which injector is operating, scroll
to
Pressure
and press
ON
.
29. Press
OVEN
, scroll to
Temp
and press
ON
.
Содержание TRACE GC Ultra