Chapter 14
Maintaining a Flame Photometric Detector
Replace the Interferential Filter
300
Maintenance and Troubleshooting Manual
8. Insert the new filter into its housing. The mirror
face must be oriented towards the flame.
9. Reassembly the photomultiplier tube and the
detector body, then fix them together tightening
the knurled nut.
10. Reconnect the signal, excitation voltage and
ignition/heating cables to the detector.
11. Switch on the GC power supply and set the
detector to the required operating conditions.
Refer to the TRACE
TM
GC Ultra
Operating
manual
for more information.