Chapter 13
Maintaining a Nitrogen Phosphorus Detector
Replace the Collecting Electrode
256
Maintenance and Troubleshooting Manual
OPERATING SEQUENCE
Replace the Collecting Electrode
Materials needed
•
Screwdriver
•
Pliers
•
Forceps or tweezers
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS
WARNING! Before replacing the collecting electrode, be sure that:
•
The flow of the detector gases has been interrupted
•
The detector has been switched off
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
for more information.