Chapter 4
Maintaining a Cold On-Column Injector for Standard Automatic Injections
Cold On-Column Injector Maintenance
82
Maintenance and Troubleshooting Manual
Cold On-Column Injector Maintenance
This injector will normally be serviced by Thermo Fisher Scientific authorized
technical personnel and do not require periodic maintenance, except for O-ring
replacement in the automatic version. The PTFE seal of the rotary valve, if worn,
can be replaced by the user.
Replace the O-ring
This sequence is specific for the Automated Operation Cold On-Column Injector
with TriPlus sampler. In this version of the Cold On-Column Injector, the o-ring
should be replaced at least every 200 injections, or every time a problem related to
O-ring damage or wear occurs. Refer to Chapter 19,
for more information. The correct sequence is explained in the Operating
Sequence
on page 83.
Possible leaks in the rotary valve
Some symptoms, particularly a low reproducibility in results, may indicate that
the connection between the rotary valve and the relevant PTFE seal is not in good
condition. Refer to Chapter 19,
To eliminate possible leaks on the rotary valve, tighten the Allen screws that fix
the upper and the lower parts of the injector, then test the injector to see if the
symptom has disappeared. If not, replace the seal of the rotary valve, as explained
in the
Operating Sequence on page 87.