Chapter 19
Results
Analytical Troubleshooting
Maintenance and Troubleshooting Manual
399
SYMPTOMS SOURCE
Results
Behavior
Characteristics
Cause
Remedy
Low reproducibility of peaks
area
Concentration not
compatible with the
dynamic range of the
detection system
Ensure that the sample
concentration is suitable for the
detection system.
Inappropriate injection
technique
Try a different injection
technique.
Injection parameters
inappropriate
Check the injection temperature
and flow rates.
Non reproducible sample
injection technique
Evaluate the sample preparation
sequences.
Compare the results with a series
of standard injections.
Leaking syringe or septum
Check and replace the syringe
and/or septum at regular
intervals.
Leaks at the injection
Check the column connections.
Run a leak check.
Poor injection technique
Carefully meter the injected
amount.
Use a clean, good-quality
syringe.
Poor split flow or ratio
control
Monitor the flow.
Replace the in-line filter.
Poor
sensitivity
with increased
retention time
Carrier gas flow rate too
low
Increase the carrier gas flow rate.
Locate and remove possible
obstructions in the carrier gas
line.
Check the injector/column
ferrules.