Chapter 16
Maintaining a Thermal Conductivity Detector
TCD Troubleshooting
344
Maintenance and Troubleshooting Manual
Symptom
Diagnosis
Remedy
Baseline drift
Faulty temperature control
Contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
Low sensitivity
Oxidized filaments
Contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
The detector does not
work.
Message:
Filament power Off
Lack of carrier, make-up or
reference gas or pressure too
low
Check the feed of the carrier, make-up
and reference gas.
Filament burnt
Contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
Table 16-2.
TCD troubleshooting table (Continued)