Chapter 1
Preventive Maintenance
Suggested Maintenance Schedule
24
Maintenance and Troubleshooting Manual
1. Only a licensed laboratory can perform this operation
Annually
Clean FID collector.
Replace PID lamp.
Regenerate the active carbon filter mounted on the carrier gas line
Regenerate the active carbon filter mounted on the split line.
Biannually
Replace FPD photomultiplier.
Clean or replace ECD source
1
As required by local law
Perform wipe test on ECD
1
Table 1-1.
Standard Maintenance Schedule