Chapter 17
Maintaining Detectors in Stack Configuration
Clean or Replace the Collecting Electrode (FPD on ECD)
364
Maintenance and Troubleshooting Manual
WARNING! Before disconnecting the cables or disassembling an FPD detector, switching off
the GC power supply is imperative to avoid electrical shocks and damages to the
instrument.
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
for more information.
1. Switch off the GC power supply and wait until
the detectors have cooled to room temperature.
2. Disconnect the signal, excitation voltage and
ignition/heating cables from the FPD detector.