Chapter 19
Analytical Troubleshooting
Peaks
396
Maintenance and Troubleshooting Manual
Behavior
Characteristics
Cause
Remedy
Fronting
Column or detector
overloaded
Decrease the injected amount
and/or analyte concentrations.
Increase the split ratio.
Column temperature too
low
Increase the temperature.
Stationary phase too thin
Use a thicker-film column.
Ghost peaks
Contaminated carrier gas
Replace the cylinder or replace
the filter.
Contamination from
laboratory glassware.
Ensure the glassware is clean and
contamination-free.
Decomposition of injected
sample
Decrease the injection port
temperature.
Use the on-column injection
technique.
Dirty injection solution
Carry out adequate clean-up of
sample prior to injection.
Ghost
peaks
Broad ghost
peaks
Contaminated inlet or
pneumatics
Remove the column and bake out
the inlet.
Use a high-quality septum.
Replace the split vent filter.
Install an in-line filter between
the pneumatics and the inlet.
Incomplete elution of
previous sample
Increase the final oven program
temperature or total run time.
Increase the column flow rate.
Irregular
shape
Chair-shaped
Solvent flooding of column
Increase the initial oven
temperature.
Reduce the injection volume
(OC).
Install a retention gap (OC).
Negative
peaks
All
Integrator wires reversed
Correct the connections.