Chapter 11
Clean the Collecting Electrode
Maintaining a FID
Maintenance and Troubleshooting Manual
203
1.
.
RIGHT DETECTOR (FID)
Flame
Off <
Base temp
300
300
Signal pA
(5.4)
Ign. tresh
2.0
Flameout retry
Off
H2
35
Off
Air
350
Off
Mkup N2
30
30
Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the FID detector,
to access the FID control table. In the following
example, a FID installed on the right channel is
considered.
2. Scroll to
Flame
and press
OFF
.
3. Scroll subsequently to
Base temp,H2
and
Air
and press
OFF
.
4. Disconnect the signal and ignition cables from
the detector.
5. Loosen the fixing nut on the front of the
detector cell and remove it.
6. Unscrew and remove the cap of the detector
cell.