Chapter 14
Replace the Heat Shields
Maintaining a Flame Photometric Detector
Maintenance and Troubleshooting Manual
307
20. Insert the mirror plug in the detector body and
fix it with the relevant Allen screws.
NOTE
Before reinserting the plug, inspect
the graphite seal inside the plug’s
housing. If the seal is damaged
and could not ensure tightness,
remove it and replace with a new
one.
21. Insert a new heat shield into its housing inside
the flange.
22. Insert the filter support into the flange and fix it
tightening the three Allen screws. Check the
O-ring that fixes the heat shield is in place
before tightening the screws.