Chapter 8
Clean or Replace the Liner
Maintaining a Purged Packed Column Injector
Maintenance and Troubleshooting Manual
137
3. Press
LEFT INLET
or
RIGHT INLET
depending on
which injector is operating. In the following
example, a PPKD injector installed on the right
channel is considered.
RIGHT INLET (PPKD)
Temp
200
200 <
Pressure
30
30
Mode:
Wide Bore
4. Scroll to
Temp
and press
OFF
.
5. When the inlet reaches the room temperature,
press
LEFT CARRIER
or
RIGHT CARRIER
depending on which injector is operating.
RIGHT CARRIER (PPKD)
Col. flow
3.0
<
Pressure
30
30
Flow mode
con pres
6. Scroll to
Pressure
and press
OFF
.
7. Unscrew the nut that retains the analytical
column.
8. Remove the column with the inserted ferrule.