Chapter 14
FPD Troubleshooting
Maintaining a Flame Photometric Detector
Maintenance and Troubleshooting Manual
313
Symptom
Diagnosis
Remedy
Low sensitivity
Low photomultiplier
excitation voltage
Set
High Voltage Mode
to
Yes
in
the FPD control table
Low sensitivity
Inadequate hydrogen flow
rate
Set the hydrogen flow rate to a proper
value. Refer to Table 14-4,
Recommended gas flow rates for FPD
on page 311
Inadequate air flow rate
Set the air flow rate to a proper value.
Refer to Table 14-4,
on page 311
and relevant note
Dirty interferential filter
Clean the interferential filter. Refer to
Clean the Interferential Filter
Operating Sequence on page 278
The mirror of the combustion
chamber is dirty
Clean the mirror. Refer to
Operating
Sequence on page 282
The flame-side heat shield is
dirty
Clean the flame-side heat shield.
Refer to
Operating Sequence on
Reduced transparency of the
heat shields
Replace the heat shields. Refer to
Operating
Sequence on page 301
Low sensitivity.
Water droplets generate
between the heat shields
Leak at the flame-side heat
shield
Replace the heat shields.
Refer to
Operating Sequence on page 301
If the symptom does not disappears,
contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
Table 14-5.
FPD troubleshooting table (Continued)
Содержание TRACE GC Ultra
Страница 4: ......
Страница 10: ...x Maintenance and Troubleshooting Manual Contents ...
Страница 20: ......
Страница 128: ......
Страница 178: ...Chapter 9 Maintaining a PTV Injector PTV Troubleshooting 178 Maintenance and Troubleshooting Manual ...
Страница 218: ......
Страница 268: ......
Страница 314: ......