Chapter 12
Wipe Test
Maintaining an Electron Capture Detector
Maintenance and Troubleshooting Manual
225
ECD Wipe Test is to be carried out under a controlled procedure as reported below:
Procedure
Wet cotton wool swab with methanol.
Wipe the outer surface of the ECD.
Remove metal exit tube from top of ECD.
Wipe the bottom of the exit tube.
Remove the two screws retaining the tower cap.
Wipe the bottom of the cylindrical surface and place the swab in a marked vial.
Repeat twice steps 1-6 with new cotton wool swabs and place in clean marked vial.
Submit the marked vials containing the swabs for the radioactivity measurement using a liquid
scintillation counter.
Notes:
The wipe test should be performed on the ECD at room temperature.
Radioactivity should be measured by a liquid scintillation spectrometer (TRI-CARB Packard or
similar) discriminating at 1-70 keV to encompass the maximum beta energy
of 67 keV of
63
Ni. The test and background samples should be measured for 10 minutes.
Suitable precautions should be taken by the person performing the test.
The test should be carried out by authorized personnel
Limit
For a source containing 370 MBq of
63
Ni, the acceptable limit is 185 Bq.