To Regenerate Carrier Gas and Split Lines Filters
Regenerating Carrier and Split Lines Filters
Maintenance and Troubleshooting Manual
389
Figure 1-1.
TRACE GC Ultra Top Cover
Removing Active Carbon Filter
1. Use the appropriate wrench the loosen each retaining nut.
CARRIER
SPLIT
PURGE
1
2
2
3
3
4
5
2
2 3
3
4
2.1. Carrier Gas Line Filter
3.2. M6x0.75 mm Retaining Nut
4.3. 1.6 mm OD Aluminium Ferrule
5.4. Arrow indicating the Flow Direction
6.5. Split Line Filter
7.
8. Finger-loosen the retaining nuts of the filter to regenerate.