Chapter 13
Maintaining a Nitrogen Phosphorus Detector
Replace the Collecting Electrode
258
Maintenance and Troubleshooting Manual
7. Remove the holder assembly of the signal cable
by unscrewing the two screws that fix it on the
detector cell.
8. Using forceps, unscrew and remove the contact
pin.
9. Using forceps or tweezers, remove the
collecting electrode through the top side of the
detector cell.
10. Place the new collecting electrode into its
housing. Ensure the insert hole is perfectly
vertically aligned to the hole present on the cell.