Chapter 4
Maintaining a Cold On-Column Injector for Standard Automatic Injections
Replace the Rotary Valve Seal
94
Maintenance and Troubleshooting Manual
22. .Adjust the position of the automatic actuator
up to align the fixing holes provided on the
actuator body with the corresponding fixing
holes located on the fixing arm of the OC head.
Fix the parts by using the two fixing screws.
23. Reconnect the ground wire coming from the
actuator motor
24. .Reconnect the cables coming from the actuator
and the TriPlus sampler to the control interface.
25. Reposition the injector/detector cover.
26. Press
LEFT CARRIER
or
RIGHT CARRIER,
depending on which injector is operating, scroll
to
Pressure
and press
ON
.
27. Switch on the TriPlus sampler.
Содержание TRACE GC Ultra
Страница 4: ......