Chapter 19
Peaks
Analytical Troubleshooting
Maintenance and Troubleshooting Manual
395
SYMPTOMS SOURCE
Peaks
Behavior
Characteristics
Cause
Remedy
Broadening
Column flow too high
Reduce the flow to slightly
above optimum.
Column flow too low
Increase the flow to slightly
above optimum.
Split flow too low in split
injection
Increase the flow to 40-50
ml/min.
Column performances
degraded
Test the column at the optimum
flow rate.
Dirty injector
Clean or replace the liner.
Stationary phase
accumulated in the outlet
Remove the last two coils from
the column.
Detector base body
temperature too low.
Increase the temperature to 5°C
below the column maximum.
Clipping
at bottom
Detector or integrator zero
set too low
Set the zero correctly.
at top
Data system zoomed in too
close
Zoom out to view the entire
chromatogram.
Detector or integrator
attenuation set too low
Set the attenuation higher.
Detector range too sensitive
Set a less-sensitive detector
range.
Incorrect input to recording
unit
Correct and check the recording
unit.
Double peaks
Injection speed too low
Inject more rapidly in a smooth
motion.
Wrong autosampler
injection speed or mode
Use a higher speed.