Chapter 2
Replace the Split Line Tubing
Maintaining a Split/Splitless Injector
Maintenance and Troubleshooting Manual
49
4. Scroll to
Temp
and press
OFF
.
5. When the inlet reaches the room temperature,
press
LEFT CARRIER
or
RIGHT CARRIER
depending on which injector is operating.
RIGHT CARRIER (S/SL)
Pressure
30.0
30.0
Col. flow
3.00
<
Lin. Veloc.
(60.9)
6. Scroll to
Pressure
and press
OFF
.
7. Remove the GC top cover.
a. (1) Lift the detector cover off the GC top
cover.
b. (2) Open the oven door and unscrew the two
top cover fastening screws.
c. (3) Push the cover back about 1 cm and lift
it up and off the GC.
8.
Split Line Tubing 1/8"
Split Line Filter
Split Line Connections
Look for the split line tubing connections.