Chapter 13
NPD Troubleshooting
Maintaining a Nitrogen Phosphorus Detector
Maintenance and Troubleshooting Manual
267
Symptom
Diagnosis
Remedy
Large negative baseline upset
at solvent elution with no
recovery to the original
baseline (“Solvent quenching
effect”)
Heating current too low.
Slightly increase the source
heating current.
Unstable baseline
Background current level too
high
Reduce the source heating
current value.
Hydrogen flow fluctuation.
Check pressure regulators on
hydrogen line.
Low carbon rejection
Hydrogen flow too high.
Reduce hydrogen flow to the
correct operative conditions.
Table 13-4.
NPD Troubleshooting Table (Continued)