Chapter 15
Replace the UV Lamp
Maintaining a Photoionization Detector
Maintenance and Troubleshooting Manual
321
4. Remove the lamp with its Viton
TM
O-ring from
the lamp holder and store it.
CAUTION
Avoid touching the lamp and
particularly the lamp window
and the O-ring with your fingers.
Use cotton gloves during this
operation.
5. Place the proper Viton
TM
O-ring on the new
lamp so to contact its flange.
6. Insert the lamp into the holder, pushing it in
until it stops.
CAUTION
While inserting the lamp, pay
attention not to damage the
contact pin inside the holder.