Chapter 11
Replace the Jet
Maintaining a FID
Maintenance and Troubleshooting Manual
199
OPERATING SEQUENCE
Replace the Jet
Materials needed
•
Screwdriver
•
Tool for removing the jet (P/N 205 019 00)
•
Forceps or tweezers
•
Jet for FID (P/N 404 043 01)
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS
1.
.
RIGHT DETECTOR (FID)
Flame
Off <
Base temp
300
300
Signal pA
(5.4)
Ign. tresh
2.0
Flameout retry
Off
H2
35
Off
Air
350
Off
Mkup N2
30
30
Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the FID detector,
to access the FID control table. In the following
example, a FID installed on the right channel is
considered.
2. Scroll to
Flame
and press
OFF
.
3. Scroll subsequently to
Base temp,H2
and
Air
and press
OFF
.