Chapter 13
Maintaining a Nitrogen Phosphorus Detector
NPD Troubleshooting
266
Maintenance and Troubleshooting Manual
Troubleshooting the NPD
Table 13-4.
NPD Troubleshooting Table
Symptom
Diagnosis
Remedy
No detector response
Source heating current too low.
Increase the heating current.
No hydrogen flow.
Turn the hydrogen flow on and
set it to a correct value.
No air flow.
Turn the air flow on and set it
to a correct value.
Source turned off
Turn the source on.
Source faulty
Replace the source
Detector response lower than
expected
Lower source temperature
Check the base body
temperature and source heating
current.
Air contamination in hydrogen
line.
Turn the source off. Increase
the hydrogen pressure for 10-
20 minutes to purge the
hydrogen line. Check the line
tightness.
High background level
Heating current too high.
Set the correct operating
parameter.
Hydrogen flow too high.
Set the correct operating
parameter.
Air and/or Make-up gas flow
too low.
Set the correct operating
parameter.
Excessive column bleed.
Column conditioning has to be
performed.
FID-like response for solvent
and other carbon based
compounds
Hydrogen flow too high.
Reduce hydrogen flow to a
lower value.