Chapter 9
Maintaining a PTV Injector
Clean or Replace the Liner
152
Maintenance and Troubleshooting Manual
3. Press
RIGHT INLET
.
RIGHT INLET (PTV)
Temp
50
50 <
Pressure
100
100
Mode
split
4. Scroll to
Temp
and press
OFF
.
5. When the inlet reaches the room temperature,
press
RIGHT CARRIER
depending on which
injector is operating.
RIGHT CARRIER (PTV)
Col. flow
2.5
<
Pressure
100
100
Flow mode
con pres
6. Scroll to
Pressure
and press
OFF
.
7. Unscrew the injector cap.
8. Remove the septum holder with septum, then
the septum support.