Chapter 7
Replacing the Septum
Maintaining a Packed Column Injector
Maintenance and Troubleshooting Manual
127
9. Insert a new septum in the septum cap.
CAUTION
Use tweezers to avoid touching
the septum with your fingers.
10. Tighten the septum cap.
CAUTION
Do not overtighten the injector
cap. You could damage the
septum and affect performance.
11. Press
LEFT CARRIER
or
RIGHT CARRIER,
depending on which injector is operating, scroll
to
Pressure
and press
ON
.
12. Press
OVEN
, scroll to
Temp
and press
ON
.
13. Press
LEFT INLET
or
RIGHT INLET,
depending on
which injector is operating, scroll to
Temp
and
press
ON
.
14. Perform a leak check according to
Sequence on page 181.