Chapter 13
Maintaining a Nitrogen Phosphorus Detector
Replace the NPD Thermionic Source
262
Maintenance and Troubleshooting Manual
6. Unscrew the grub screw on the collar of the
source holder.
7. Remove the four retaining screws of the source
metallic socket and pull out the source.
8. Gently replace the thermionic source, pushing it
in straight.
9. Fix the thermionic source into its housing,
tightening the grub screw.