Chapter 2
Clean or Replace the Liner
Maintaining a Split/Splitless Injector
Maintenance and Troubleshooting Manual
39
18. Clean the septum support from possible
fragments left by the septum and reinsert it into
the injector
19. Place the septum holder on the top of the
septum support.
20. Tighten the injector cap to finger tight.
CAUTION
Do not overtighten the injector
cap. You could damage the
septum and affect performance.
21. Press
LEFT CARRIER
or
RIGHT CARRIER,
depending on which injector is operating, scroll
to
Pressure
and press
ON
.
22. Press
OVEN
, scroll to
Temp
and press
ON
.
23. Press
LEFT INLET
or
RIGHT INLET,
depending on
which injector is operating, scroll to
Temp
and
press
ON
.
24. Perform a leak check according to
Sequence on page 181.