Chapter 9
Adjust a Quartz Wool Packing
Maintaining a PTV Injector
Maintenance and Troubleshooting Manual
169
15. Tighten the liner cap using the screwdriver
provided with the GC.
16. Remove the septum from the septum holder
using a non-metallic sharp tool.
17. Insert a new septum into the septum holder.
CAUTION
Use tweezers to avoid touching
the septum with your fingers.