Chapter 17
Maintaining Detectors in Stack Configuration
Clean or Replace the Collecting Electrode (FID on ECD)
350
Maintenance and Troubleshooting Manual
1.
.
AUXILIARY
Signal
<
Detector
Temperature Zones
Pressure Channel
Press
AUX
.
2. Scroll to
Detector
and press
ENTER
.
3.
.
AUX DETECTOR (FID)
Flame
Off <
Base temp
300
Off
Signal pA
(5.4)
Ign. tresh
2.0
Flameout retry
Off
H2
35
Off
Air
350
Off
In the
AUX DETECTOR (FID)
table, scroll to
Flame
and press
OFF
.
4. Scroll subsequently to
Base temp,H2
and
Air
and press
OFF
.
5. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the ECD, to access
the ECD control table.
.
XXX DETECTOR (ECD)
Base temp
250
Off <
ECD temp
250
Off
Ref current nA
1.0
Freq kHz
(2.15)
Pulse amp V
25
Pulse width us
1.0
Mkup (N2)
40
Off
6. Scroll subsequently to
Base temp
and
ECD
temp
and press
OFF
.
7. Scroll to
Mkup(N2)
and press
OFF
.