Chapter 5
Replace the Rotary Valve Seal
Maintaining a Cold On-Column Injector Direct Narrow-Bore
Maintenance and Troubleshooting Manual
101
OPERATING SEQUENCE
Replace the Rotary Valve Seal
Materials needed
•
5.5 mm hexagonal wrench
•
PTFE seal (P/N 29007106)
•
3 mm Allen wrench
•
4-mm Allen wrench
•
5.5-mm wrench
1. While in stand-by condition, press
OVEN
to
access the oven control table.
.
OVEN
Temp
70
70 <
Initial time
1.0
Ramp 1
20.0
2. Scroll to
Temp
and press
OFF
.
3. Press
LEFT CARRIER
or
RIGHT CARRIER
depending on which injector is operating, scroll
to
Pressure
and press
OFF
.
LEFT CARRIER (OCI)
Pressure
30.0
30.0
Col. flow
3.00
<
Lin. Veloc.
(60.9)
4. Remove the injector/detector cover.
5. Switch off the TriPlus sampler
Содержание TRACE GC Ultra