Chapter 12
Maintaining an Electron Capture Detector
Replace the Silver Seal
232
Maintenance and Troubleshooting Manual
OPERATING SEQUENCE
Replace the Silver Seal
Materials needed
•
Silver seal (P/N 290 150 58)
•
ECD fixing tool (P/N 205 021 50)
•
Tweezers
WARNING! Any maintenance or repair operations involving the radioactive source must be
performed ONLY by qualified personnel duly authorized to handle radioactive
material. Please contact your local Thermo Fisher Scientific Technical Service office.
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS
1. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the ECD, to access
the ECD control table. In the following
example, an ECD installed on the right channel
is considered.
.
RIGHT DETECTOR (ECD)
Base temp
250
Off <
ECD temp
250
Off
Ref current nA
1.0
Freq kHz
(2.15)
Pulse amp V
25
Pulse width us
1.0
Mkup (N2)
40
Off
2. Scroll subsequently to
Base temp
and
ECD
temp
and press
OFF
.
3. Scroll to
Mkup(N2)
and press
OFF
.