Chapter 19
Baseline
Analytical Troubleshooting
Maintenance and Troubleshooting Manual
393
Behavior
Characteristics
Cause
Remedy
High-standing current
Excessive column
stationary phase bleeding
Check the oven temperature,
ensuring it doesn't exceed the
column upper limit.
Recondition the column.
Replace the column.
Loose connections
Ensure all interconnections and
screw connections are tight.
Irregular
shape
Dip after solvent
peak
Detector contaminated
Bake out or clean the detector.
S-shaped
Excessive column bleed
during column temperature
programming
Reduce the upper column
temperature. Bake out the
column. Install a high
temperature column.
Oxygen contamination is
decomposing the stationary
phase
Install oxygen filters in carrier
gas line.Check pneumatic and
inlet systems for leaks. Use
correct gas purity with low
oxygen content.
Square waves
Large AC power
fluctuations; heavy
equipment on same line
Use a dedicated clean AC line of
sufficient amperage
Noise
High-frequency
noise
Contaminated detector
Isolate the detector from
electronics. If noise disappears,
clean the collector.
Combustion-gas flow too
low or too high
Check the detector-gas flows
Contaminated column
Condition the column.
Contaminated detector gas
supply
Check the gas purity and install
appropriate filters
Defective electrometer
Replace electrometer.