Chapter 4
Maintaining a Cold On-Column Injector for Standard Automatic Injections
Replace the O-Ring
84
Maintenance and Troubleshooting Manual
5. Unscrew the seal holder and remove it.
6. Remove the O-ring from the seal holder with
the aid of a sharp, non-metallic tool.
7. Insert a new O-ring in the relevant seat.
CAUTION
Use tweezers to avoid touching
the O-ring with your fingers.