Chapter 3
Clean or Replace the Liner
Maintaining a Large Volume Splitless Injector
Maintenance and Troubleshooting Manual
63
9. Use the liner cap wrench provided with the GC
to remove the injector liner cap.
10. Use tweezers to remove the liner with the
graphite seal.
CAUTION
Be careful not to break the glass
liner inside the injector when
removing it. Glass splinters
might fall into the lower part of
the vaporization chamber.
If the liner breaks, refer to the
Operating Sequence on
page 67.
11. If you are going to use a new liner, go to step
14. If you are going to clean the liner, put the
dirty liner into an ultrasonic bath filled with a
methanol/acetone mixture (1:1) and sonicate it
for about half an hour.