Chapter 13
Replace the Collecting Electrode
Maintaining a Nitrogen Phosphorus Detector
Maintenance and Troubleshooting Manual
257
1. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the NPD detector,
to access the NPD control table. In the
following example, a NPD installed on the right
channel is considered.
.
RIGHT DETECTOR (NPD)
Source cur,A
Off <
Base temp
300
Off
Signal pA
(10.2)
Polarizer V
3.5
H2 delay time
Off
H2
2.3
Off
Air
70
Off
Mkup N2
15
Off
2. Scroll to
Base temp
and press
OFF
.
3. Scroll subsequently to
Source curr,H2
,
Air
and
Mkup
and press
OFF
.
4. Disconnect the signal and source cables from
the detector.
5. Loosen the fixing screw on the front of the
detector cell and remove it.
6. Unscrew and remove the cap of the detector
cell.