Chapter 19
Analytical Troubleshooting
Baseline
394
Maintenance and Troubleshooting Manual
Behavior
Characteristics
Cause
Remedy
Noise
High-frequency
noise
Detector temperature higher
than column maximum
temperature
Reduce the detector temperature
to the column temperature upper
limit.
External electrical
interference
Attach an AC line monitor and
check purity of the AC supply
Loose column fittings
Tighten fittings accordingly.
Loose detector electrical
connections
Make sure the leads are properly
connected.
Spiking
Defective electrometer or
amplifier
Replace the electrometer or
amplifier
F.s. column too close to
flame (FID)
Lower the column to the correct
position (2-3 mm below the tip
of the jet).
Dirty jet or detector
Isolate the detector from the
electronics. If noise disappears,
clean the jet and collector.
External electrical
interference
Attach an AC line monitor and
check the purity of the AC
supply
FID temperature too low
Increase the temperature to at
least 150°C