Chapter 9
Maintaining a PTV Injector
PTV Troubleshooting
176
Maintenance and Troubleshooting Manual
Table 9-4.
PTVLV Troubleshooting Table
Symptom
Diagnosis
Remedy
Poor reproducibility
of results
Transfer temperature too low
Set
Transfer temp
to a higher
value.
Transfer time too low
Set
transfer time
or
cleaning
time
to a value close to the whole GC
run time.
Injection speed set too high
Reduce injection speed.
Injected volume too large
Reduce the amount of sample to be
injected.
Initial temperature set too low Set
Inject temperature
to an
higher value.
Injection pressure set too high Set
Inject pressure
to a lower
value.
Lack of an adequate quartz
wool packing inside the liner
Fix the quartz wool packing inside the
liner. Refer to
page 165. Replace the liner with another
filled with quartz wool.
Operating Sequence on page 151.
Discrimination of
volatile compounds
Injection speed set too low
Set
Inject speed
to a higher value.
Initial temperature set too
high
Set
Inject temperature
to a lower
value.
Split flow set too high
Lower the split flow.
Lack of an adequate quartz
wool packing inside the liner
Fix the quartz wool packing inside the
liner. Refer to
page 165.
Replace the liner with another filled
with quartz wool.
Operating Sequence on page 151.