Chapter 16
Maintaining a Thermal Conductivity Detector
TCD Troubleshooting
342
Maintenance and Troubleshooting Manual
Troubleshooting the TCD
Table 16-2.
TCD troubleshooting table
Symptom
Diagnosis
Remedy
Baseline fluctuations
Unstable regulation of the
flow rate of the gases
Check the controllers of the carrier,
reference and make-up gases work
well.
Leaks on the gas lines
Check the connections of the carrier,
reference and make-up gas are tight.
Inlet pressure of gases set too
low
Set the pressure of the carrier,
reference and make-up gas to a proper
value.
Refer to the TRACE
TM
GC Ultra
Operating Manual
for instructions.
Trans temp
set too high
Set
Trans temp
to a proper value
(about 10 to 20°C below the
Block
temp
value).
Faulty temperature regulation
Contact your customer support
organization.
Refer to Appendix B,
, for contact
information.
Baseline drift
Unstable regulation of the
flow rate of the gases
Check the controllers of the carrier,
reference and make-up gases work
well.
Leaks on the gas lines
Check the connections of the carrier,
reference and make-up gas are tight.
Column conditioning not
correctly performed
Recondition the column according to
the manufacturer’s instructions.
Septum bleeding
Check the operating temperature of
the septum is adequate.
Replace the septum if necessary.