Chapter 2
Clean or Replace the Liner
Maintaining a Split/Splitless Injector
Maintenance and Troubleshooting Manual
35
3. Press
LEFT INLET
or
RIGHT INLET
depending on
which injector is operating. In the following
example, a S/SL injector installed on the right
channel is considered.
RIGHT INLET (S/SL)
Temp
250
250 <
Pressure
100
100
Mode
split
4. Scroll to
Temp
and press
OFF
.
5. When the inlet reaches the room temperature,
press
LEFT CARRIER
or
RIGHT CARRIER
depending on which injector is operating.
RIGHT CARRIER (S/SL)
Pressure
30.0
30.0
Col. flow
3.00
<
Lin. Veloc.
(60.9)
6. Scroll to
Pressure
and press
OFF
.
7. Unscrew the injector cap.
8. Remove the septum holder with septum, then
the septum support.