Chapter 11
Clean the Jet
Maintaining a FID
Maintenance and Troubleshooting Manual
197
6. Using the tool provided with the GC, unscrew
the jet.
7. Mechanically remove any contaminating
material and sweep with clean compressed air if
necessary.
8. Handling the jet with forceps or tweezers, rinse
the jet with distilled water, then with methanol.
9. Place the jet on a paper towel and let it air dry.
PRECAUTIONS
10. When the jet is dry, place the jet into the
housing and tighten it with the proper tool.
Ensure the jet is perfectly vertically aligned to
avoid damaging its ceramic part.