Chapter 12
Clean the Collecting Electrode
Maintaining an Electron Capture Detector
Maintenance and Troubleshooting Manual
227
1. Press
LEFT DETECTOR
or
RIGHT DETECTOR
,
depending on the location of the ECD, to access
the ECD control table. In the following
example, an ECD installed on the right channel
is considered.
.
RIGHT DETECTOR (ECD)
Base temp
250
Off <
ECD temp
250
Off
Ref current nA
1.0
Freq kHz
(2.15)
Pulse amp V
25
Pulse width us
1.0
Mkup (N2)
40
Off
2. Scroll subsequently to
Base temp
and
ECD
temp
and press
OFF
.
3. Scroll to
Mkup(N2)
and press
OFF
.
4. Disconnect the signal, excitation and heater
cables from the detector.
5. Remove the cover and the detector chimney
unscrewing the two retaining Allen screws. Use
a 1.5 mm Allen wrench.
6. Unscrew the collecting electrode and remove it
using tweezers.
7. Clean the collecting electrode, washing it with
toluene and hexane, preferably in an ultrasonic
bath. In the most critical cases fine emery paper
can be used.