Chapter 10
Ensuring Tightness
Performing a Leaks Check (Capillary/Packed/UFM Columns)
186
Maintenance and Troubleshooting Manual
1. This item appears if the relevant channel is present and configured.
8. Scroll to the channel of interest and press
ENTER
to open the following menu:
L. COL. LEAK CHECK
Start Leak Check
Leak Check Settings
.
9. Scroll to
Leak Check Settings
and press
ENTER
to open the relevant menu.
L. COL. SETT.
Check press 200
Sensitivity 5.0
10. In
Check press
line set the pressure (pressurization value) at which the leak
check must be performed. The range is comprised from 10 to 999 kPa
(1.45-145 psi).
11. In
Sensitivity
line set the maximum value the pressure can drop during the
test. The range is comprised from 1 to 10 kPa (0.145-1.45 psi).
12. Press
CLEAR
to exit
Leak Check Settings
menu and return to the previous
menu.
L. COL. LEAK CHECK
Start leak check
Leak Check settings
13. Select the
Start leak check
command to start the operation. The split and
purge valves of the selected channel are automatically closed and the channel
is pressurized with carrier gas to the leak check set point.
CHECKING L COLUMN
Pressure (200)
Elapsed time 0.90
Use <STOP> to abort