Chapter 2
Maintaining a Split/Splitless Injector
Replace a Broken Liner
40
Maintenance and Troubleshooting Manual
OPERATING SEQUENCE
Replace a Broken Liner
Materials needed
•
Liner cap removal tool (P/N 205 070 10)
•
Glass liner (see Table 2-2 on page 29)
•
Tweezers
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the injector must be cooled to room temperature.
PRECAUTIONS
1. While in stand-by condition, press
OVEN
to
access the oven control table.
.
OVEN
Temp
50
50 <
Initial time
2
Ramp 1
10.0
2. Scroll to
Temp
and press
OFF
.
3. Press
LEFT INLET
or
RIGHT INLET
depending on
which injector is operating. In the following
example, a S/SL injector installed on the right
channel is considered.
RIGHT INLET (S/SL)
Temp
250
250 <
Pressure
100
100
Mode
split
4. Scroll to
Temp
and press
OFF
.